Evaluation of prediction errors in nine intraocular lens calculation formulas using an explainable machine learning model
BMC Ophthalmology volume 24, Article number: 531 (2024)
Abstract
Background
The purpose of the study was to evaluate the relationship between prediction errors (PEs) and ocular biometric variables in cataract surgery using nine intraocular lens (IOL) formulas with an explainable machine learning model.
Methods
We retrospectively analyzed the medical records of consecutive patients who underwent standard cataract surgery with a Tecnis 1-piece IOL (ZCB00) at a single center. We calculated predicted refraction using the following IOL formulas: Barrett Universal II (BUII), Cooke K6, EVO V2.0, Haigis, Hoffer QST, Holladay 1, Kane, SRK/T, and PEARL-DGS. We used a LightGBM-based machine learning model to evaluate the explanatory power of ocular biometric variables for PEs and assessed the relationship between PEs and ocular biometric variables using Shapley additive explanation (SHAP) values.
Results
We included 1,430 eyes of 1,430 patients in the analysis. The SRK/T formula exhibited the highest R2 value (0.231) in the test set among the machine-learning models. In contrast, the Kane formula exhibited the lowest R2 value (0.021) in the test set, indicating that the model could explain only 2.1% of the PEs using ocular biometric variables. BUII, Cooke K6, EVO V2.0, Haigis, Hoffer QST, Holladay 1, PEARL-DGS formulas exhibited R2 values of 0.046, 0.025, 0.037, 0.194, 0.106, 0.191, and 0.058, respectively. Lower R2 values for the IOL formulas corresponded to smaller SHAP values.
Conclusion
The explanatory power of currently used ocular biometric variables for PEs in new-generation formulas such as BUII, Cooke K6, EVO V2.0 and Kane is low, implying that these formulas are already optimized. Therefore, the introduction of new ocular biometric variables into IOL calculation formulas could potentially reduce PEs, enhancing the accuracy of surgical outcomes.
Significance
Previous studies have evaluated the accuracy of intraocular lens power formulas for specific subgroups of patients according to the ocular biometric variables and provided valuable insights into the strengths and weaknesses of various formulas. However, the ocular biometric parameters showed significant associations with each other.
AbstractSection What this study addsCompared with those of formally generated formulas, recent intraocular lens formulas yielded relatively stable outcomes across a wide range of ocular biometric variables. Ocular biometric variables have no significant effect on postoperative prediction errors.
Background
The primary goal of modern cataract surgery is to achieve optimal vision and predicted refraction. The prediction accuracy of postoperative refraction has increased significantly in recent years. The development of advanced optical technologies and intraocular lens (IOL) power calculation formulas have improved refractive outcomes. Conventional IOL formulas, such as the SRK/T [1], Haigis [2], and Barrett Universal II (BUII) [3] formulas, are based on theoretical lens calculations. More complex optics and effective lens position calculation for IOL power calculation have been incorporated in recently developed formulas, such as EVO 2.0 and Kane formulas [4].
Several studies have attempted to develop optimal IOL formulas for specific eye groups by classifying the eyes according to the axial length (AL) and comparing the refractive outcomes among the subgroups [5,6,7,8,9]. Other studies have revealed remarkable variations among the refractive outcomes of different subgroups according to the anterior chamber depth (ACD, measured from corneal epithelium to lens) [10, 11]. New-generation formulas demonstrate higher overall accuracy by incorporating more ocular biometric variables [12, 13]. Despite this advancement, there remains a need to further enhance IOL formulas for improved precision.
Ocular biometric variables show significant correlations with each other [14]. Thus, multiple ocular biometric factors affect the refractive outcomes of cataract surgery. Multivariable analysis must be conducted considering these parameters simultaneously to calculate the prediction errors (PEs), rather than attributing the PEs to individual biometric parameters.
In this study, we used a machine learning model to predict the refractive outcomes of nine IOL power formulas using ocular biometric variables to investigate the overall influence of ocular biometric variables on PE and compare the prediction outcomes of the formulas.
Methods
This study was approved by the Institutional Review Board of the Seoul National University Hospital (SNUH; IRB No. 2112-132-1284) and adhered to the principles of the Declaration of Helsinki. The Institutional Review Board waived the requirement for obtaining written informed consent owing to the retrospective study design and anonymization of patient information.
Study population
We retrospectively reviewed the medical records of patients who had undergone standard cataract surgery between August 1, 2018, and December 31, 2021. The inclusion criteria were as follows: (1) operation using Tecnis ZCB00 (Johnson & Johnson Vision Care, Inc., Santa Ana, CA, USA) IOL insertion in the bag, and (2) age of at least 19 years old. The exclusion criteria were as follows: (1) history of previous vitrectomy, corneal refractive surgery, or other corneal operation; (2) incidence of severe intraoperative or postoperative complication, such as zonular dialysis, posterior capsular rupture, or the use of capsular tension ring or iris retractor; (3) combined operation for the correction of glaucoma, pterygium, or vitrectomy; (4) cataract operation with large corneal or limbal incision or limbal relaxing incision; (5) absence of postoperative manifest refraction data; (6) postoperative best-corrected visual acuity (BCVA) of worse than < 20/40; (7) failure of ocular biometric examination; (8) inability to calculate IOL power owing to extreme refractive target. The first eye that underwent surgery was included if both eyes were eligible for inclusion.
Preoperative ophthalmic evaluation
In accordance with the SNUH preoperative cataract examination protocol, all patients underwent a comprehensive ophthalmologic examination, which included BCVA assessment, slit-lamp biomicroscopy, dilated funduscopic examination, ocular biometric measurement (IOLMaster 700; Carl Zeiss, Germany), autokeratometry (KR-7100; Topcon, Japan), anterior segment topographic measurements (Orbscan II; Bausch and Lomb, USA), specular microscopy (NSP-9900; Konan Medical, Japan), optical coherence tomography (Heidelberg Spectralis; Heidelberg Engineering, Germany), and ultra-widefield fundus photography (Optos California; Optos, USA), preoperatively.
Surgical procedures
All cataract surgeries were performed by the experienced surgeons at SNUH, as described below. The procedures were as follows: 2.2-mm or 2.75-mm small, clear corneal incision, continuous curvilinear capsulorrhexis, phacoemulsification of the crystalline lens, and implantation of the IOL in the capsular bag. The corneal incision was made at a temporal or superior location at the discretion of the surgeon.
Prediction of postoperative refractive errors
We predicted the postoperative refractive errors using nine IOL power calculation formulas: Barrett Universal II (BUII), Cooke K6, EVO V2.0, Haigis, Hoffer QST, Holladay 1, Kane, SRK/T, and PEARL-DGS. We used the manufacturers’ and IOL calculators’ recommended constants available from online (https://www.iolcon.org/lensesTable.php; accessed April 5, 2024) for each IOL formula. The A constant of the IOL used was 119.3. The Haigis constants for a0, a1, and a2 were − 1.302, 0.210, and 0.251, respectively. The SRK/T formula requires the A constant for the IOL, corneal power, and AL. An investigator (RO) manually entered the data into online calculators for the BUII, Cooke K6, EVO V2.0, Hoffer-QST, Kane, and PEARL-DGS formulas (https://calc.apacrs.org/barrett_universal2105/, https://cookeformula.com/, https://www.evoiolcalculator.com/calculator.aspx, https://hofferqst.com/, https://www.iolformula.com/, and https://iolsolver.com/main, respectively). Another investigator (CHY) evaluated the results to determine the plausibility.
Postoperative refractive outcome analysis
All patients underwent routine postoperative examinations. We assessed the manifest refraction 1 month postoperatively. We adhered to the previously proposed protocols for the postoperative refractive outcome analysis [15, 16], and defined PE as the difference between the spherical equivalent of the postoperative manifest refraction and formula PE using the IOL power implanted. To eliminate systematic error, we zeroed out the mean PE by adjusting the PE for each eye up or down by an amount equal to the mean PE. Negative and positive PEs indicate myopic and hyperopic outcomes, respectively [16]. We defined the absolute prediction error (APE) as the absolute value of PE and calculated the mean value of PE (ME), median of value APE (MedAE), mean value of APE (MAE), and percentages of eyes within ± 0.25 diopter (D), ± 0.50 D, ± 0.75 D, and ± 1.00 D from the target refraction.
Development of machine learning models to estimate PEs after cataract surgery
We used LightGBM, which was developed by Microsoft and Peking University, to evaluate the relative influence of ocular biometric parameters on the PEs of the nine IOL formulas [17]. LightGBM is one of the most popular machine learning models owing to its superior accuracy, computational speed, and memory consumption compared with those of other machine-learning models [17]. LightGBM was trained using the lightgbm library version 3.3.2, with the following hyperparameters: n_estimators = 10,000, learning_rate = 0.01, max_depth = 8, and otherwise, with the default value. We subsequently divided the dataset into development and test sets at an 8:2 ratio and used five-fold cross-validation in the development process. We divided the development set into training and validation sets of 80% and 20%, respectively, for each fold. As a result, the training, validation, and test sets were exclusively constructed at the patient level. We fitted and validated the LightGBM model using the training and validation sets and estimated the PE using the following variables: age, gender, and ocular biometric measurements (including AL, ACD, mean keratometry [K], LT, central corneal thickness [CCT], and horizontal corneal diameter [CD]). We developed five models for the test set to estimate PE for each IOL formula. The predicted values of the five models were averaged for each test sample to determine the final PE using the formula.
Model performance evaluation and SHAP
We calculated the R squared value (R2, coefficient of determination), defined as the proportion of the variation in the dependent variable that can be predicted from the independent variable 1 - RSS/TSS (where RSS is the sum of squares of residuals, and TSS is the total sum of squares), to measure the predictive performance of the models. This value is a measure of how well observed outcomes are replicated by the model based on the proportion of the total variation of outcomes explained by the model. A value of 1 indicates that the model predicts 100% of the relationship, whereas a value of 0.5 indicates that the model predicts 50% of the relationship [18]. We used the bootstrap method to calculate the 95% confidence intervals (CIs) of R2. From the test set, the same amount of data as the test set was resampled, with allowance for repetitive samples for the MAE and R2 evaluation. This process was repeated 10,000 times to calculate CIs.
We used the SHAP method, a game-theoretic technique used to explain the output of machine learning models, to interpret the model [19]. SHAP values yield quantified contributions, thereby intuitively demonstrating the effect of each feature in terms of the shift of the model output from the base value. SHAP values quantify the effect of individual parameters on the model output and estimated PE. Further details on the SHAP method have been described in the article by Lundberg and Lee [19]. We calculated the SHAP value by determining the average change relative to the presence or absence of individual features after constructing a model with several features. The SHAP value of each feature is an indicator of its strength in terms of positive or negative prediction of the model. A larger absolute SHAP value indicates a greater effect of the feature on the prediction of the model. We calculated the SHAP values to determine the contribution of each variable and their correlation with the PEs of the formulas. Features with positive signs indicate a positive effect on PEs, whereas those with negative signs indicate a negative effect on PEs. The partial dependence plot (PDP) presents the marginal effect of features on the predicted outcome of a machine learning model.
Development process and analysis
We used Python ver. 3.7.11 (https://www.python.org), scikit-learn library ver. 1.0.2, and shap library ver. 0.41.0 to develop and analyze the performance of the model. An investigator (RO) performed all developments and inferences on a private server equipped with a central processing unit (CPU) with 32 GB of RAM and an NVIDIA GeForce GTX 3090 with a 24 GB graphics processing unit (GPU; Nvidia). We used scipy library ver. 1.7.3 and scikit_posthocs library ver. 0.7.0 for statistical analysis. In addition, we used Student’s t-test, Friedman test, post hoc pairwise Wilcoxon test with Holm’s adjustment, Cochrane Q test, and post hoc pairwise Dunn’s test with Holm’s adjustment for comparisons.
Results
Among the 3,188 eyes of 2,269 patients with cataracts, 1,430 eyes of 1,430 patients (mean age: 69.71 ± 9.14 years, 898 (62.8%) females) were included in this study. Table 1 summarizes the demographic and clinical characteristics of the patients. Table 2 summarizes the predictive performance of the IOL power formulas for postoperative refractive outcomes after zeroing adjustment. The difference in the APE values among the formulas was statistically significant (P < 0.05, Friedman test). Supplementary Table 1 presents the results of the pairwise comparison. The Cooke K6 formula exhibited the lowest MAE (0.325) and BUII formula exhibited the lowest MedAE (0.254) values, whereas the SRK/T and Holladay 1 formulas exhibited the highest MAE (0.378) and MedAE (0.307) values, respectively. However, no statistical differences were observed between the Kane, BUII, and Cooke K6 formulas in terms of the MAE value. The Cochrane Q test revealed significant differences between the formulas in terms of the percentages of eyes within the given error range (P < 0.001 for all ranges). The Kane formula exhibited the highest percentage of eyes in the given error ranges. Supplementary Table 2 presents the statistical results for the percentage of eyes within the given error range using post hoc Dunn’s test with Holm’s adjustment.
LightGBM models were constructed to estimate the PEs for each IOL formula. Table 3 summarizes the R2 values of the ensemble models for the estimation of PEs using the IOL formulas in the test set. The R2 values ranged from 0.021 to 0.231. Notably, the SRK/T and Kane formulas exhibited the highest and lowest R2 values in the test set, respectively (R2 value: 0.231 and 0.021, respectively). For BUII, Cooke K6, EVO V2.0, and Kane formula, the R2 values for the models were not significantly different from zero.
The SHAP values are described in Fig. 1. The colored dots represent the SHAP value of each eye. The x-axis represents the SHAP values, wherein the negative values correspond to the eyes contributing to a negative PE and positive values correspond to the eyes contributing to a positive PE. SHAP values of < 0.0 and > 0.0 indicate myopic and hyperopic shifts, respectively. The colors ranging from blue to red represent the values of ocular biometric variables. The red dots represent eyes with greater values, whereas the blue dots represent the eyes with smaller values. The average absolute SHAP values for each variable indicate the contribution of the variation in the variable within the formula. The variable at the top represents the most important variable in the estimation of the PE for each formula. The importance of the variables decreases in descending order.
Shapley Additive Explanations (SHAP) values and mean absolute SHAP values for the IOL formulas. ACD, anterior chamber depth measured from corneal epithelium to lens; AL, axial length; BUII, Barrett Universal II; CCT, central corneal thickness; CD, horizontal corneal diameter; IOL, intraocular lens; K, Keratometry; LT, lens thickness. The Shapley additive explanation (SHAP) values and mean absolute SHAP values for each intraocular lens (IOL) formula. The SHAP value for each participant is visually presented as a colored dot. The x-axis represents the SHAP values, wherein negative values correspond to the eyes contributing to a negative prediction error (PE) and positive values correspond to a positive PE. Colors ranging from blue to red denote the value of the ocular biometric variables, with red dots representing eyes with greater values and blue dots representing eyes with smaller values. The average absolute SHAP values for each variable indicate the contribution of the variable variation within the formula. The height of the horizontal bar indicates the contribution of the variable, with larger bars signifying higher contributions and smaller bars indicating lower contributions
A lower R2 value of the IOL formula indicated a smaller absolute SHAP value. Figure 2 presents the PDPs between the SHAP values and ocular biometric variables. The x- and y-axes represent the value of the variable and the SHAP value, respectively. The SHAP values in the figures represent the marginal impact of the variables on the PEs based on the presence or absence of other variables. A longer AL was independently associated with more myopic PE in the SRK/T formula. The remaining formulas showed irregular trends. A steeper K results in myopic PE in the SRK/T formula, whereas it leads to hyperopic PEs in the Haigis formula. All variables exhibited a nearly flat curve in the PDP plots for the Kane formula, indicating that the performance efficiency of the formula was less affected by these variables.
Partial dependence plots (PDPs) between the Shapley Additive Explanations (SHAP) values and ocular biometric variables. ACD, anterior chamber depth measured from corneal epithelium to lens; AL, axial length; BUII, Barrett Universal II; CCT, central corneal thickness; CD, horizontal corneal diameter; IOL, intraocular lens; K, Keratometry; LT, lens thickness. The partial dependence plots (PDPs) between SHAP values and ocular biometric variables present the marginal impact of each variable on each intraocular lens (IOL) formula. The x- and y-axes represent the value of the variable and the SHAP value, respectively. The blue dots represent the eyes. SHAP values of < 0.0 and > 0.0 result in myopic and hyperopic shifts, respectively
Discussion
We developed machine learning models to predict PEs in cataract surgery in this study. The R2 values were generally low, ranging from 0.021 to 0.232. The SRK/T formula exhibited the highest R2 value for estimating PE, whereas the BUII, Cooke K6, EVO V2.0, and Kane formula exhibited the lowest R2 value, which did not statistically differ from zero. These findings indicate that ocular biometric parameters, including AL, ACD, K, LT, CCT, and CD, have significant effects on the PE of the SRK/T formula. However, they had no significant effect on the PE of the BUII, Cooke K6, EVO V2.0, and Kane formula. To the best of our knowledge, this is the first study to use ocular biometric variables to estimate PEs using machine learning.
Previous studies have evaluated the accuracy of IOL power formulas for specific subgroups of patients according to the ocular biometric variables [6, 7, 20,21,22,23,24,25,26,27,28] and provided valuable insights into the strengths and weaknesses of various formulas. The Kane formula exhibits a significantly lower MAE value and higher accuracy in eyes with long AL. Moreover, it outperforms other formulas, including the Hill-RBF 2.0 and BUII formulas, in eyes with short AL [12, 29]. Several studies have used other ocular biometric parameters in combination with AL. Kim et al. categorized eyes based on the AL, K, and ACD values and compared the accuracies of the Haigis, Hoffer Q, Holladay 1, SRK/T, and BUII formulas in each subgroup [14]. Hipólito-Fernandes et al. revealed that the Kane, PEARL-DGS, and EVO V.2.0 formulas yielded reliable and stable results in subgroups with extreme combinations of ACD and LT [30]. Although previous studies have provided valuable insights into ocular biometric combinations, an optimal formula for each subgroup of eyes remains to be established.
The ocular biometric parameters AL, K, ACD, LT, WTW, and CCT showed significant associations with each other. Kim et al. conducted a large-scale ocular biometric analysis and revealed that AL was positively correlated with ACD, WTW, and CCT and negatively correlated with K and LT [14]. Another larger study revealed similar tendencies among AL, ACD, K, and LT [31]. Furthermore, these parameters also show associations with age [14, 31,32,33,34]. Thus, subgroup analysis using individual variables introduced significant confounding factors, given the inherent correlations among the ocular biometric parameters. The effects of these confounding factors cannot be adequately addressed in studies using such designs. Multivariate analysis must be conducted to predict PEs and identify the best-fitting formula owing to the interdependence of these variables.
We incorporated nine ocular biometric parameters to estimate PEs using machine learning models. Traditional regression models consider all relationships as linear, thereby limiting their ability to evaluate complex interactions between variables. In contrast, machine learning models incorporate nonlinear relationships, thereby enhancing their predictive capabilities. However, the mechanisms underlying machine learning models are challenging owing to their complex structures. The SHAP method addresses this limitation by aiding in the interpretation of the outcomes of machine learning models. Moreover, it enables the exploration of the importance and dependence of variables.
The machine learning models produced R² scores below 30% for all formulas, suggesting that at least 70% of the variability cannot be explained by the ocular biometric variables included in this study. The machine learning model exhibited the lowest R2 score of 0.021 for the Kane formula. The machine learning models for BUII, Cooke K6, EVO V2.0, and Kane formula failed to introduce significant performance in estimating the variability in the predicted PE, indicating their stability across ocular biometric parameter ranges. In contrast, the R2 scores of the third-generation formulas, SRK/T, Haigis, and Holladay 1, were relatively high, indicating a stronger association between the predicted PEs and ocular biometric variables.
The impact of each variable on the prediction of PE can be discerned (Fig. 1). Notably, the absolute SHAP values of the BUII, Cooke K6, EVO V2.0, and Kane formulas were lower than those of the Haigis, Holladay 1, and SRK/T formulas. The PDP plots illustrate the adjusted influence of each variable while considering the influence of the other variables. Thus, these graphs depict the response of different formulas to changes in ocular biometry. An increase in AL leads to the SRK/T formula predicting a negative PE, whereas smaller K, larger ACD, and LT values lead to the SRK/T formula predicting greater PEs. In contrast, LT and K, rather than AL, exhibit significant effects on the Haigis formula.
The findings of our study highlight different aspects of the objectives and designs of numerous previous studies that attempted to differentiate subgroups based on the range of ocular biometric parameters and compared various formulas within these subgroups to identify the best formula in terms of PE or APE. However, the outcomes varied among research groups. Moreover, the differences often failed to demonstrate statistical differences in subgroup analyses. The findings of our study can be used to understand these divergent findings. New-generation formulas, such as the BUII, Cooke K6, EVO V2.0, and Kane, consistently yielded stable results across a wide range of parameters. In contrast, the SRK/T, Holladay 1, and Haigis formulas exhibited a less consistent performance in the subgroup analysis, which may be attributed to the PEs being explainable using known ocular biometric variables.
Our study demonstrated that the new-generation formulas are generally stable across a wide range of known ocular biometric parameters. We suggest that further research should focus not on the previous known ocular biometric parameters but on the new variables which can have affect on the PEs, such as the lens vault, angle kappa, and accuracy of the ocular biometric measurements [23, 35, 36]. New variables using the mathematical combination of the known variables, such as AL/corneal radius [37], might have a role in the change of the PEs.
This study has some limitations. First, our sample size was relatively small for the machine learning model. However, a sample size exceeding 1000 is sufficient to demonstrate this tendency. Further large-scale studies could reveal a higher proportion of potentially predictable variance. Second, we only included a single type of IOL in this study; thus, the findings of our study must be validated before being applied to other IOLs. Lastly, the machine learning model for each formula could be overfitted to our dataset, which could have resulted in bias. Moreover, comparisons have been made between these overfitted models, but not with an ideal prediction algorithm.
In conclusion, machine learning algorithms can predict a portion of PEs; however, they have limitations in explaining the variability of PEs using ocular biometric parameters. Newer IOL formulas, such as the BUII, Cooke K6, EVO V2.0, and Kane formulas, demonstrated relatively stable outcomes across a wide range of ocular biometric variables, indicating that these formulas are already optimized by ocular biometric variables used in this study. Introducing additional ocular biometric variables beyond those currently used in the IOL calculation formula may improve the accuracy of future surgical results.
Data availability
The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.
Abbreviations
- ACD:
-
anterior chamber depth measured from corneal epithelium to anterior lens surface
- AL:
-
axial length
- APE:
-
absolute prediction error
- CCT:
-
central corneal thickness
- CD:
-
horizontal corneal diameter
- D:
-
diopter
- DM:
-
diabetes mellitus
- HTN:
-
hypertension
- IOL:
-
intraocular lens
- K:
-
keratometry
- LT:
-
lens thickness
- ME:
-
mean prediction error
- MAE:
-
mean absolute prediction error
- MedAE:
-
median absolute prediction error
- PE:
-
prediction error
- SD:
-
standard deviation
- SE:
-
spherical equivalent
References
Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:528 and 1993;19(5):444–446.
Haigis W, Hoffer KJ et al. The Haigis Formula. In: Aramberri J, eds. Intraocular Lens Calculations. Springer Nature 2024:625–36.
Barrett GD. An improved universal theoretical formula for intraocular lens power prediction. J Cataract Refract Surg. 1993;19:713–20.
Connell BJ, Kane JX. Comparison of the Kane formula with existing formulas for intraocular lens power selection. BMJ open Ophthalmol. 2019;4:e000251.
Day AC, Foster PJ, Stevens JD. Accuracy of intraocular lens power calculations in eyes with axial length < 22.00 mm. Clin Exp Ophthalmol. 2012;40:855–62.
Abulafia A, Barrett GD, Rotenberg M, et al. Intraocular lens power calculation for eyes with an axial length greater than 26.0 mm: comparison of formulas and methods. J Cataract Refract Surg. 2015;41:548–56.
Moschos MM, Chatziralli IP, Koutsandrea C. Intraocular lens power calculation in eyes with short axial length. Indian J Ophthalmol. 2014;62:692.
Wang J-K, Hu C-Y, Chang S-W. Intraocular lens power calculation using the IOLMaster and various formulas in eyes with long axial length. J Cataract Refract Surg. 2008;34:262–7.
Yoon JH, Whang W-J. Comparison of accuracy of six modern intraocular Lens Power calculation formulas. Korean J Ophthalmol. 2023;37:380–6.
Gökce SE, Montes De Oca I, Cooke DL, et al. Accuracy of 8 intraocular lens calculation formulas in relation to anterior chamber depth in patients with normal axial lengths. J Cataract Refract Surg. 2018;44:362–8.
Melles RB, Holladay JT, Chang WJ. Accuracy of intraocular Lens calculation formulas. Ophthalmology. 2018;125:169–78.
Darcy K, Gunn D, Tavassoli S, et al. Assessment of the accuracy of new and updated intraocular lens power calculation formulas in 10 930 eyes from the UK National Health Service. J Cataract Refract Surg. 2020;46:2–7.
Savini G, Di Maita M, Hoffer KJ, et al. Comparison of 13 formulas for IOL power calculation with measurements from partial coherence interferometry. Br J Ophthalmol. 2021;105:484–9.
Kim S, Oh R, Kim MK, et al. SS-OCT–based ocular biometry in an adult Korean population with cataract. J Cataract Refract Surg. 2023;49:453–9.
Hoffer KJ, Savini G. Update on intraocular lens power calculation study protocols: the better way to design and report clinical trials. Ophthalmology. 2021;128:e115–20.
Wang L, Koch DD, Hill W, et al. Pursuing perfection in intraocular lens calculations: III. Criteria for analyzing outcomes. J Cataract Refract Surg. 2017;43:999–1002.
Ke G, Meng Q, Finley T et al. Lightgbm: a highly efficient gradient boosting decision tree. Adv Neural Inf Process Syst. 2017;30.
Fox RA, Steel RGD, Torrie JH. Principles and procedures of statistics with special reference to the Biological sciences. Inc Stat. 1961;11. https://doiorg.publicaciones.saludcastillayleon.es/10.2307/2987461.
Lundberg SM, Lee S-I. A unified approach to interpreting model predictions. Adv Neural Inf Process Syst. 2017;30.
Eom Y, Kang S-Y, Song JS, et al. Comparison of Hoffer Q and Haigis formulae for intraocular lens power calculation according to the anterior chamber depth in short eyes. Am J Ophthalmol. 2014;157:818–24.
Mo E, Lin L, Wang J, et al. Clinical accuracy of 6 intraocular lens power calculation formulas in elongated eyes, according to anterior chamber depth. Am J Ophthalmol. 2022;233:153–62.
Faramarzi A, Aghajani A, Ghiasian L. Accuracy of various intraocular lens power calculation formulas in steep corneas. J Ophthalmic Vis Res. 2017;12:385.
Yan C, Yao K. Effect of lens vault on the accuracy of intraocular lens calculation formulas in shallow anterior chamber eyes. Am J Ophthalmol. 2022;233:57–67.
Lee Y, Kim MK, Oh JY, et al. Intraocular lens power calculation in eyes with a shallow anterior chamber depth and normal axial length. PLoS ONE. 2023;18:e0288554.
Gavin EA, Hammond CJ. Intraocular lens power calculation in short eyes. Eye. 2008;22:935–8.
26 Németh J, Fekete O, Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg. 2003;29:85–8.
Wang J-K, Chang S-W. Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths. Int J Ophthalmol. 2013;6:150.
Reitblat O, Levy A, Kleinmann G, et al. Intraocular lens power calculation for eyes with high and low average keratometry readings: comparison between various formulas. J Cataract Refract Surg. 2017;43:1149–56.
Melles RB, Kane JX, Olsen T, et al. Update on intraocular lens calculation formulas. Ophthalmology. 2019;126:1334–5.
Hipólito-Fernandes D, LuÃs ME, Serras-Pereira R, et al. Anterior chamber depth, lens thickness and intraocular lens calculation formula accuracy: nine formulas comparison. Br J Ophthalmol. 2022;106:349–55.
Lei Q, Tu H, Feng X, et al. Distribution of ocular biometric parameters and optimal model of anterior chamber depth regression in 28,709 adult cataract patients in China using swept-source optical biometry. BMC Ophthalmol. 2021;21:1–8.
Fotedar R, Wang JJ, Burlutsky G, et al. Distribution of axial length and ocular biometry measured using partial coherence laser interferometry (IOL Master) in an older white population. Ophthalmology. 2010;117:417–23.
Ferreira TB, Hoffer KJ, Ribeiro F, et al. Ocular biometric measurements in cataract surgery candidates in Portugal. PLoS ONE. 2017;12:e0184837.
Lim LS, Saw S-M, Jeganathan VSE, et al. Distribution and determinants of ocular biometric parameters in an Asian population: the Singapore malay eye study. Invest Ophthalmol Vis Sci. 2010;51:103–9.
Langenbucher A, Hoffmann P, Cayless A, et al. Impact of uncertainties in biometric parameters on intraocular lens power formula predicted refraction using a Monte-Carlo simulation. Acta Ophthalmol. 2024;102:e285–95.
Marques JH, Baptista PM, Ribeiro B, et al. Intraocular lens power calculation: angle κ and ocular biomechanics. J Cataract Refract Surg. 2024;50:345–51.
Omoto MK, Torii H, Hayashi K, et al. Ratio of axial length to corneal Radius in Japanese patients and accuracy of Intraocular Lens Power Calculation Based on Biometric Data. Am J Ophthalmol. 2020;218:320–9.
Acknowledgements
Not applicable.
Funding
None.
Author information
Authors and Affiliations
Contributions
RO acquired and analyzed the data, drafted the initial manuscript, and revised the manuscript. JYO, HJC, MKK conceptualized and designed the study, and collected data. CHY conceptualized and designed the study, coordinated and supervised data collection, analyzed the data, critically reviewed the manuscript, and revised the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study was approved by the Institutional Review Board of Seoul National University Hospital (IRB No. 2112-132-1284) and was conducted in accordance with the Declaration of Helsinki. The need for written informed consent was waived because of the retrospective design and the use of deidentified patient data.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Meeting presentation
16th Joint Meeting of Korea-China-Japan Ophthalmologists, 2023.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Oh, R., Oh, J.Y., Choi, H.J. et al. Evaluation of prediction errors in nine intraocular lens calculation formulas using an explainable machine learning model. BMC Ophthalmol 24, 531 (2024). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12886-024-03801-2
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12886-024-03801-2